RSRS-V Transformer capacity and no-load/load tester
RSRS-V can test the transformer capacity, type, no-load current, no-load loss, short circuit (load) loss, resistance voltage, etc. It is small in volume, light in weight, in high accuracy, with steady property, easy to operation can replace the traditional testers. It is displayed by LCD, operated by menu, with friendly interface, is the first choice for all customers.
Product Features
1. Measure the transformer no-load current, no-load loss; short-circuit voltage, short circuit (load) loss, capacity.
2. Test methods: three-table.
3. The device can auto check if the wire was correctly connected, and display the vector diagram of three phase current, voltage when test the three phase transformer.
4. Single machine can make the loading test for substation transformer under 1600KVA.
5. All the related test results are automatically corrected.
6. 320x240 LCD, friendly man-machine interface, brilliance control, adapts to all seasons.
7. With electronic clock, auto record the data, time, it is good for saving and managing the test results.
8. Panel type printer can print the test results on site.
1. Service conditions | ||
Temperature | -5°C~40°C | |
Humidity | <95%(25°C) | |
Altitude | <2500m | |
Outside interference | no exceptionally strong vibration, no extra - strong electromagnetic fields. | |
Power supply | 220VAC±10%,45Hz~55Hz. | |
2. Main technical parameters | ||
Active parts: | ||
Current output | 0 ~ 10A | |
Voltage measurement | 0 ~ 10V | |
Test Capacity | 10KV oil-immersed, dry-type transformers: 30KVA ~ 2500KVA | |
35KV oil-immersed transformers: 50KVA ~ 31500KVA | ||
20KV dry type transformers: 50KVA ~ 2500KVA | ||
35KV dry type transformers: 50KVA ~ 2000KVA | ||
Features parts: | ||
Voltage | 0~450V (phase voltage), 0 ~ 800V (line voltage), auto-switch the measurement range | |
Current | 0 ~ 100A, auto-switch the measurement range | |
Frequency | 45Hz ~ 65Hz | |
3. Test accuracy | ||
Voltage | 20 ~ 100V, ± 0.2% FS; 100 ~ 450V, ± 0.2% | |
Current | 0.5 ~ 10A, ± 0.2% FS; 10 ~ 100A, ± 0.2% | |
Power | ± 0.5% (CosΦ> 0.1), ± 1.0% (0.02 <CosΦ <0.1) | |
4. Dielectric strength | The insulation resistance between voltage, current input terminal and the chassis≥100MΩ. | |
Between power supply input terminal and the shell can withstand 2KV (rms), for 1 minute | ||
5. Dimension | 400×300×170mm | |
6. Weight | 6kg (not including test lines) |